Shopping Cart

No products in the cart.

BS EN IEC 61223-3-6:2020

$198.66

Evaluation and routine testing in medical imaging departments – Acceptance and constancy tests. Imaging performance of mammographic X-ray equipment used in a mammographic tomosynthesis mode of operation

Published By Publication Date Number of Pages
BSI 2020 60
Guaranteed Safe Checkout
Categories: ,

If you have any questions, feel free to reach out to our online customer service team by clicking on the bottom right corner. We’re here to assist you 24/7.
Email:[email protected]

IEC 61223-3-6:2020 applies to the performance of MAMMOGRAPHIC X-RAY EQUIPMENT when used in MAMMOGRAPHIC TOMOSYNTHESIS modes of operation, with respect to image quality and dose. Excluded from the scope of this document are: – MAMMOGRAPHIC X-RAY EQUIPMENT modes of operation other than MAMMOGRAPHIC TOMOSYNTHESIS; – 2D images synthesised from the tomosynthesis images; – reconstructive TOMOGRAPHY other than MAMMOGRAPHIC TOMOSYNTHESIS; – CT SCANNERS covered by IEC 61223-3-5. IEC 61223-3-6:2020 defines a) the essential parameters which describe the acceptability criteria of MAMMOGRAPHIC TOMOSYNTHESIS modes of operation of MAMMOGRAPHIC X-RAY EQUIPMENT with regard to image quality and dose, b) the methods of testing whether measured quantities related to those parameters comply with specified tolerances, and c) CONSTANCY TEST frequency when required. This document is intended to be applied along with the acceptability criteria included in IEC 61223-3-2 or equivalent protocol for 2D mammography which are also relevant for MAMMOGRAPHIC TOMOSYNTHESIS modes of operation. These methods mainly rely on non-invasive measurements that use appropriate test equipment and are performed during or after the installation. Signed statements covering steps in the installation procedure can be used as part of the ACCEPTANCE TEST. Tests required by a higher level of compliance take precedence over similar tests with a lower level of compliance. When the results of the ACCEPTANCE TEST are in compliance with the expected values, the BASELINE VALUES for the subsequent CONSTANCY TESTS are established.

PDF Catalog

PDF Pages PDF Title
2 undefined
5 Annex ZA(normative)Normative references to international publicationswith their corresponding European publications
7 English
CONTENTS
11 FOREWORD
13 INTRODUCTION
14 1 Scope and object
2 Normative references
15 3 Terms, definitions, symbols and abbreviated terms
3.1 Terms and definitions
18 3.2 Symbols and abbreviated terms
4 General aspects of the acceptance test
4.1 Levels of requirements
4.1.1 Local regulatory
4.1.2 Contractual
4.1.3 General
4.2 General conditions in test procedures
19 4.3 Documents and data for the tests
4.4 Test conditions
20 4.5 Scope of tests
4.6 Test equipment
4.6.1 General
21 4.6.2 Analysis software
4.6.3 Dosimeter
4.7 Evaluating the test results
22 5 General aspects of constancy tests
5.1 Establishment of baseline values
5.2 Frequency of constancy tests
6 Summary of tests for mammographic tomosynthesis equipment
Tables
Table 1 – Tests, test frequencies, and test objects used in this document
23 7 Inventory and initial tests for mammographic tomosynthesis equipment
7.1 Requirements
24 7.2 Test method
7.3 Constancy testing
7.3.1 Test method
7.3.2 Frequency of testing
7.4 Action to be taken
8 Alignment and collimation checks
8.1 Requirements
8.2 Test method
25 8.3 Constancy testing
8.3.1 Test method
8.3.2 Frequency of testing
8.4 Equipment
8.5 Action to be taken
9 AEC-system
9.1 General
Figures
Figure 1 – Set-up for measuring the alignment between the reconstructed and the irradiated volume at the chest wall edge of the patient support
26 9.2 Short term reproducibility
9.2.1 Requirements
9.2.2 Test method
9.2.3 Constancy testing
9.2.4 Equipment
9.2.5 Action to be taken
9.3 Long term reproducibility
9.3.1 Requirements
27 9.3.2 Test method
9.3.3 Constancy testing
9.3.4 Action to be taken
9.4 AEC performance
9.4.1 Requirements
9.4.2 Test method
28 Figure 2 – Top and 3D view of setup for the AEC performance measurements
29 Figure 3 – Placement of ROI for the AEC performance measurement
Table 2 – Height of the compression paddle when using different PMMA thicknesses
30 9.4.3 Constancy testing
9.4.4 Equipment
9.4.5 Action to be taken
10 Image receptor
10.1 Response function
10.1.1 General
31 10.1.2 Requirements
10.1.3 Test method
10.1.4 Constancy testing
10.1.5 Action to be taken
32 10.2 Detector element failure
10.2.1 Requirements
10.2.2 Test method
10.2.3 Constancy testing
10.2.4 Equipment
10.2.5 Action to be taken
10.3 Uncorrected defective detector elements
10.3.1 General
10.3.2 Requirements
10.3.3 Test method
33 10.3.4 Constancy testing
10.3.5 Equipment
10.3.6 Action to be taken
10.4 System projection MTF
10.4.1 General
10.4.2 Requirements
34 10.4.3 Test method
10.4.4 Constancy testing
10.4.5 Equipment
10.4.6 Action to be taken
11 Image quality of the reconstructed image
11.1 Phantom testing
11.1.1 General
11.1.2 Requirements
35 11.1.3 Test method
11.1.4 Constancy testing
11.1.5 Action to be taken
11.2 z-resolution (artefact spread function)
11.2.1 Requirements
11.2.2 Test method
36 Figure 4 – Top and 3D view of setup for the evaluation of z-resolution
37 11.2.3 Constancy testing
11.2.4 Equipment
11.2.5 Action to be taken
12 Missed tissue
12.1 General
Figure 5 – Front and side view of setup for the evaluation of z-resolution
38 12.2 Missed tissue at chest wall side in the reconstructed tomosynthesis volume
12.2.1 Requirements
12.2.2 Test method
12.2.3 Constancy testing
12.2.4 Equipment
12.2.5 Action to be taken
12.3 Missed tissue at the top and bottom of the reconstructed tomosynthesis volume
12.3.1 Requirements
12.3.2 Test method
39 12.3.3 Constancy testing
Figure 6 – Configuration for the determination of missed tissue for curved paddles
40 12.3.4 Equipment
12.3.5 Action to be taken
13 Artefacts in the tomosynthesis data sets
13.1 General
13.2 Artefact evaluation
13.2.1 Requirements
13.2.2 Test method
13.2.3 Constancy testing
13.2.4 Equipment
13.2.5 Action to be taken
13.3 Geometric distortion
13.3.1 Requirements
41 13.3.2 Test method
Figure 7 – Top and 3D view of setup for the evaluation of geometric distortion
42 13.3.3 Equipment
13.3.4 Action to be taken
14 Dosimetry for digital breast tomosynthesis
14.1 Requirements
Figure 8 – Front and side view of setup for the evaluation of geometric distortion
43 14.2 Test method
Table 3 – Limits for AGD versus the thickness of the PMMA and the height of the compression paddle
44 14.3 Constancy testing
14.3.1 Test method
14.3.2 Frequency of testing
14.4 Equipment
14.5 Action to be taken
Figure 9 –Top and 3D view of position of dosimeter to determine the incident air kerma for dose estimation
45 Annex A (informative)Tables for dosimetry calculation in digital breast tomosynthesis
Table A.1 – g factors for breasts simulated with PMMA
Table A.2 – c factors for breasts simulated with PMMA
46 Table A.3 – Typical HVL measurements for different tube voltage and target filter combinations
Table A.4 – s factors for clinically used spectra
Table A.5 – s factors for clinically used spectra with W target material
47 Table A.6 – s factors for a tungsten target filtered by 0,5 mm aluminium
Table A.7 – s factors for a tungsten target filtered by 0,7 mm aluminium
48 Table A.8 – T factors vs. PMMA thickness for a variety of scan angles
49 Annex B (normative)Guidance on action to be taken
51 Annex C (informative)Image quality evaluation
52 Annex D (informative)Artefacts
53 Bibliography
57 Index of defined terms
BS EN IEC 61223-3-6:2020
$198.66