Shopping Cart

No products in the cart.

BS EN 62209-1:2016 – TC:2020 Edition

$280.87

Tracked Changes. Measurement procedure for the assessment of specific absorption rate of human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices – Devices used next to the ear (Frequency range of 300 MHz to 6 GHz)

Published By Publication Date Number of Pages
BSI 2020 509
Guaranteed Safe Checkout
Categories: ,

If you have any questions, feel free to reach out to our online customer service team by clicking on the bottom right corner. We’re here to assist you 24/7.
Email:[email protected]

IEC 62209-1:2016 specifies protocols and test procedures for measurement of the peak spatial-average SAR induced inside a simplified model of the head with defined reproducibility. It applies to certain electromagnetic field (EMF) transmitting devices that are positioned next to the ear, where the radiating structures of the device are in close proximity to the human head, such as mobile phones, cordless phones, certain headsets, etc. These protocols and test procedures provide a conservative estimate with limited uncertainty for the peak-spatial SAR that would occur in the head for a significant majority of people during normal use of these devices. The applicable frequency range is from 300 MHz to 6 GHz. This second edition cancels and replaces the first edition published in 2005. This edition constitutes a technical revision.

PDF Catalog

PDF Pages PDF Title
283 English
CONTENTS
292 FOREWORD
294 INTRODUCTION
295 1 Scope
2 Normative references
3 Terms and definitions
300 4 Symbols and abbreviations
4.1 Physical quantities
301 4.2 Constants
4.3 Abbreviations
5 Measurement system specifications
5.1 General requirements
303 5.2 Phantom specifications (shell and liquid)
304 5.3 Hand and device holder considerations
5.4 Scanning system requirements
5.5 Device holder specifications
305 5.6 Characteristics of the readout electronics
6 Protocol for SAR assessment
6.1 General
6.2 Measurement preparation
6.2.1 Preparation of tissue-equivalent liquid and system check
306 6.2.2 Preparation of the wireless device under test (DUT)
307 6.2.3 Operating modes
308 6.2.4 Positioning of the DUT in relation to the phantom
310 Figures
Figure 1 – Vertical and horizontal reference lines and reference Points A, B on two example device types: a full touch screen smart phone (top) and a keyboard handset (bottom)
313 Figure 2 – Cheek position of the wireless device on the left side of SAM where the device shall be maintained for the phantom test set-up.
Figure 3 – Tilt position of the wireless device on the left side of SAM
314 Figure 4 – An alternative form factor DUT and standard coordinateand reference points applied
315 6.2.5 Test frequencies for DUT
6.3 Tests to be performed
317 6.4 Measurement procedure
6.4.1 General
Figure 5 – Block diagram of the tests to be performed
318 6.4.2 General procedure
319 Tables
Table 1 – Area scan parameters
Table 2 – Zoom scan parameters
320 6.4.3 SAR measurements of handsets with multiple antennas or multiple transmitters
Figure 6 – Orientation of the probe with respect to the line normal to the phantom surface, shown at two different locations
324 Table 3 – Example method to determine the combined SAR value using Alternative 1
326 6.5 Post-processing of SAR measurement data
6.5.1 Interpolation
Figure 7 – Measurement procedure for different types of correlated signals
327 6.5.2 Extrapolation
6.5.3 Definition of the averaging volume
6.5.4 Searching for the maxima
6.6 Fast SAR testing
6.6.1 General
328 6.6.2 Fast SAR measurement procedure A
330 6.6.3 Fast SAR testing of required frequency bands
331 6.6.4 Fast SAR measurement procedure B
333 6.7 SAR test reduction
6.7.1 General requirements
Figure 8 – The Fast SAR measurement procedure B.
334 6.7.2 Test reduction for different operating modes in the same frequency band using the same wireless technology
335 6.7.3 Test reduction based on characteristics of DUT design
336 6.7.4 Test reduction based on SAR level analysis
337 Table 4 – Threshold values TH(f) used in this proposed test reduction protocol
338 6.7.5 Test reduction based on simultaneous multi-band transmission considerations
Figure 9 – Modified chart of 6.4.2
339 7 Uncertainty estimation
7.1 General considerations
7.1.1 Concept of uncertainty estimation
340 7.1.2 Type A and Type B evaluation
7.1.3 Degrees of freedom and coverage factor
341 7.2 Components contributing to uncertainty
7.2.1 General
7.2.2 Calibration of the SAR probes
346 7.2.3 Contribution of mechanical constraints
347 7.2.4 Phantom shell
348 7.2.5 Device positioning and holder uncertainties
350 7.2.6 Tissue-equivalent liquid parameter uncertainty
352 Table 5 – Example uncertainty template and example numerical values for dielectric constant () and conductivity (() measurement
353 7.2.7 Uncertainty in SAR correction for deviations in permittivity and conductivity
354 Table 6 –Uncertainty of Formula (41) as a function of the maximum change in permittivity or conductivity
355 7.2.8 Measured SAR drift
356 7.2.9 RF ambient conditions
357 7.2.10 Contribution of post-processing
358 Table 7 – Parameters for the reference function f1 in Formula (48)
362 7.2.11 SAR scaling uncertainty
Figure 10 – Orientation and surface of the averaging volume relative to the phantom surface
363 7.2.12 Deviation of experimental sources
7.2.13 Other uncertainty contributions when using system validation sources
Table 8 – Uncertainties relating to the deviations of theparameters of the standard waveguide source from theory
364 7.3 Calculation of the uncertainty budget
7.3.1 Combined and expanded uncertainties
7.3.2 Maximum expanded uncertainty
Table 9 – Other uncertainty contributions relating to the dipole sources described in Annex G.
Table 10 – Other uncertainty contributions relating to the standard waveguide sources described in Annex G
366 Table 11 – Example of measurement uncertainty evaluation template for handset SAR test
369 Table 12 – Example of measurement uncertainty evaluation template for system validation
371 Table 13 – Example of measurement repeatability evaluation template for system check (applicable for one system).
373 7.4 Uncertainty of fast SAR methods based on specific measurement procedures and post-processing techniques
7.4.1 General
7.4.2 Measurement uncertainty evaluation
378 Table 14 – Measurement uncertainty budget for relative fast SAR tests
380 Table 15 – Measurement uncertainty budget for system check using fast SAR methods
382 8 Measurement report
8.1 General
8.2 Items to be recorded in the measurement report
385 Annexes
Annex A (normative) Phantom specifications
A.1 Rationale for the SAM phantom shape
A.2 SAM phantom specifications
386 Figure A.1 – Illustration of dimensions in Table A.1 and Table A.2
387 Table A.1 – Dimensions used in deriving SAM phantom from the ARMY 90th percentile male head data (Gordon et al. [56])
Table A.2 – Additional SAM dimensions compared with selected dimensions from the ARMY 90th-percentile male head data (Gordon et al. [56]) – specialist head measurement section
388 Figure A.2 – Close-up side view of phantom showing the ear region
Figure A.3 – Side view of the phantom showing relevant markings
390 Figure A.4 – Sagittally bisected phantom with extended perimeter (shown placed on its side as used for device SAR tests)
Figure A.5 – Picture of the phantom showing the central strip
391 A.3 Flat phantom specifications
Figure A.6 – Cross-sectional view of SAM at the reference plane
392 A.4 Tissue-equivalent liquids
Figure A.7 – Dimensions of the elliptical phantom
393 Table A.3 – Dielectric properties of the head tissue-equivalent liquid
394 Annex B (normative) Calibration and characterization of dosimetric probes
B.1 Introductory remarks
395 B.2 Linearity
B.3 Assessment of the sensitivity of the dipole sensors
B.3.1 General
B.3.2 Two-step calibration procedures
397 Table B.1 – Uncertainty analysis for transfer calibration using temperature probes
399 Figure B.1 – Experimental set-up for assessment of the sensitivity (conversion factor) using a vertically-oriented rectangular waveguide
400 Table B.2 – Guidelines for designing calibration waveguides
401 B.3.3 One step calibration procedures
Table B.3 – Uncertainty analysis of the probe calibration in waveguide
402 Figure B.2 – Illustration of the antenna gain evaluation set-up
403 Table B.4 – Uncertainty template for evaluation of reference antenna gain
404 Table B.5 – Uncertainty template for calibration using reference antenna
405 B.3.4 Coaxial calorimeter method
406 Figure B.3 – Schematic of the coaxial calorimeter system
407 B.4 Isotropy
B.4.1 Axial isotropy
B.4.2 Hemispherical isotropy
Table B.6 – Uncertainty components for probe calibration using thermal methods
408 Figure B.4 – Set-up to assess spherical isotropy deviation in tissue-equivalent liquid
409 Figure B.5 – Alternative set-up to assess spherical isotropy deviation in tissue-equivalent liquid
410 Figure B.6 – Experimental set-up for the hemispherical isotropy assessment
Figure B.7 – Conventions for dipole position (ξ) and polarization (θ )
411 Figure B.8 – Measurement of hemispherical isotropy with reference antenna
412 B.5 Lower detection limit
B.6 Boundary effects
B.7 Response time
413 Annex C (normative) Post-processing techniques
C.1 Extrapolation and interpolation schemes
C.1.1 Introductory remarks
C.1.2 Interpolation schemes
C.1.3 Extrapolation schemes
C.2 Averaging scheme and maximum finding
C.2.1 Volume average schemes
C.2.2 Extrude method of averaging
414 C.2.3 Maximum peak SAR finding and uncertainty estimation
C.3 Example implementation of parameters for scanning and data evaluation
C.3.1 General
C.3.2 Area scan measurement requirements
C.3.3 Zoom scan
Figure C.1 – Extrude method of averaging
415 C.3.4 Extrapolation
C.3.5 Interpolation
C.3.6 Integration
Figure C.2 – Extrapolation of SAR data to the inner surface of the phantom based on a fourth-order least-square polynomial fit of the measured data (squares)
416 Annex D (normative) SAR measurement system verification
D.1 Overview
D.2 System check
D.2.1 Purpose
417 D.2.2 Phantom set-up
D.2.3 System check source
418 D.2.4 System check source input power measurement
Figure D.1 – Test set-up for the system check
419 D.2.5 System check procedure
420 D.3 System validation
D.3.1 Purpose
D.3.2 Phantom set-up
D.3.3 System validation sources
421 D.3.4 Reference dipole input power measurement
D.3.5 System validation procedure
422 D.3.6 Numerical target SAR values
423 Table D.1 – Numerical target SAR values (W/kg) for standard dipole and flat phantom
424 Table D.2 – Numerical target SAR values for waveguides specified in Clause G.2 placed in contact with flat phantom [94]
425 D.4 Fast SAR method system validation and system check
D.4.1 General
D.4.2 Fast SAR method system validation
426 D.4.3 Fast SAR method system check
427 Annex E (normative) Interlaboratory comparisons
E.1 Purpose
E.2 Phantom set-up
E.3 Reference wireless handsets
E.4 Power set-up
428 E.5 Interlaboratory comparison – Procedure
429 Annex F (informative) Definition of a phantom coordinate system and a device under test coordinate system
Figure F.1 – Example reference coordinate system for the left ERP of the SAM phantom
430 Figure F.2 – Example coordinate system on the device under test
431 Annex G (informative) SAR system validation sources
G.1 Standard dipole source
Table G.1 – Mechanical dimensions of the reference dipoles
432 G.2 Standard waveguide source
Figure G.1 – Mechanical details of the standard dipole
433 Figure G.2 – Standard waveguide source (dimensions are according to Table G.2)
Table G.2 – Mechanical dimensions of the standard waveguide
434 Annex H (informative) Flat phantom
435 Figure H.1 – Dimensions of the flat phantom set-up used for deriving the minimal phantom dimensions for W and L for a given phantom depth D
Figure H.2 – FDTD predicted uncertainty in the 10 g peak spatial-average SAR as a function of the dimensions of the flat phantom compared with an infinite flat phantom, at 800 MHz
436 Table H.1 – Parameters used for calculation of reference SAR values in Table D.1
437 Annex I (informative) Example recipes for phantom head tissue-equivalent liquids
I.1 Overview
I.2 Ingredients
438 I.3 Tissue-equivalent liquid formulas (permittivity/conductivity)
Table I.1 – Suggested recipes for achieving target dielectric parameters: 300 MHz to 900 MHz
439 Table I.2 – Suggested recipes for achieving target dielectric parameters: 1 450 MHz to 2 000 MHz
440 Table I.3 – Suggested recipes for achieving target dielectric parameters: 2 100 MHz to 5 800 MHz
441 Annex J (informative) Measurement of the dielectric properties of liquidsand uncertainty estimation
J.1 Introductory remarks
J.2 Measurement techniques
J.2.1 General
J.2.2 Instrumentation
J.2.3 General principles
442 J.3 Slotted coaxial transmission line
J.3.1 General
J.3.2 Equipment set-up
J.3.3 Measurement procedure
Figure J.1 – Slotted line set-up
443 J.4 Contact coaxial probe
J.4.1 General
J.4.2 Equipment set-up
444 Figure J.2 – An open-ended coaxial probe with innerand outer radii a and b, respectively
445 J.4.3 Measurement procedure
J.5 TEM transmission line
J.5.1 General
J.5.2 Equipment set-up
446 J.5.3 Measurement procedure
Figure J.3 – TEM line dielectric test set-up [143]
447 J.6 Dielectric properties of reference liquids
448 Table J.1 – Parameters for calculating the dielectric properties of various reference liquids
Table J.2 – Dielectric properties of reference liquids at 20 °C
450 Annex K (informative) Measurement uncertainty of specific fast SAR methods and fast SAR examples
K.1 General
K.2 Measurement uncertainty evaluation
K.2.1 General
451 K.2.2 Probe calibration and system calibration drift
K.2.3 Isotropy
452 K.2.4 Sensor positioning uncertainty
K.2.5 Sensor location sensitivity
453 K.2.6 Mutual sensor coupling
K.2.7 Sensor coupling with the DUT
K.2.8 Measurement system immunity / secondary reception
K.2.9 Deviations in phantom shape
454 K.2.10 Spatial variation in dielectric parameters
455 Table K.1 – Measurement uncertainty budget for relative fast SAR tests complying with Annex K requirements, for tests performed within one frequency band and modulation
457 Table K.2 – Measurement uncertainty budget for system check using fast SAR methods complying with Annex K requirements
459 K.3 Fast SAR examples
K.3.1 General
Figure K.1 – SAR values for twelve hypothetical test configurations measured in the same frequency band and modulation (e.g. GSM 900 MHz) using a hypothetical full SAR (full SAR) and two fast SAR (fast SAR 1 and fast SAR 2) evaluations
460 K.3.2 Example 1: Tests for one frequency band and mode
Table K.3 – Measurements conducted according to Step a)
461 Table K.4 – Measurements conducted according to Step b)
Table K.5 – Measurements conducted according to Step c)
462 Table K.6 – Measurements conducted according to 6.4.2, Step 2)
463 Table K.7 – Measurements conducted according to 6.4.2, Step 3)
Table K.8 – Measurements conducted according to 6.4.2, Step 4)
464 K.3.3 Example 2: Tests over multiple frequency bands and modes
Table K.9 – Fast SAR measurements conducted according to Step a)
465 Table K.10 – Fast SAR measurements showing highest SAR value according to Step b)
Table K.11 – Full SAR measurements conducted according to Step b)
466 Table K.12 – Fast SAR measurements showing values according-to requirements in Step c)
Table K.13 – Full SAR measurements conducted according to Step c)
467 K.3.4 Example 3: Tests for one frequency band and mode (Procedure B)
Table K.14 – Fast SAR measurements showing values accordingto requirements in Step e)
Table K.15 – Full SAR measurements conducted according to Step e)
468 Table K.16 – Measurements conducted according to Step a)
469 Table K.17 – Measurements conducted according to Step b)
Table K.18 – Measurements conducted according to Step c)
470 Table K.19 – Measurements conducted according to Step e)
471 K.3.5 Example 4: Tests over multiple frequency bands and modes (Procedure B)
Table K.20 – Measurements conducted according to Step f)
472 Table K.21 – Fast SAR measurements conducted according to Step a)
Table K.22 – Full SAR measurements conducted according to Step b)
473 Table K.23 – Full SAR measurements conducted according to Step e)
474 Table K.24 – Full SAR measurements conducted according to Step e)
475 Annex L (informative) SAR test reduction supporting information
L.1 General
L.2 Test reduction based on characteristics of DUT design
L.2.1 General
L.2.2 Statistical analysis overview
476 L.2.3 Analysis results
Figure L.1 – Distribution of “Tilt/Cheek”
Table L.1 – The number of handsets used for the statistical study
477 Table L.2 – Statistical analysis results of P(Tilt/Cheek > x) for various x values
Table L.3 – Statistical analysis results of P(Tilt/Cheek > x)for 1 g and 10 g peak spatial-average SAR
478 Table L.4 – Statistical analysis results of P(Tilt/Cheek > x)for various antenna locations
Table L.5 – Statistical analysis results of P(Tilt/Cheek > x) for various frequency bands
479 L.2.4 Conclusions
L.2.5 Expansion to multi transmission antennas
L.2.6 Test reduction based on analysis of SAR results on other signal modulations
Table L.6 – Statistical analysis results of P(Tilt/Cheek > x) for various device types
481 L.3 Test reduction based on SAR level analysis
L.3.1 General
Figure L.2 – SAR relative to SAR in position with maximum SAR in GSM mode
482 L.3.2 Statistical analysis
Figure L.3 – Two points identifying the minimum distance between the position of the interpolated maximum SAR and the points at 0,6 ( SARmax
483 Figure L.4 – Histogram for Dmin in the case of GSM 900 and iso-level at 0,6 × SARmax
Table L.7 – Distance Dmin* for various iso-level values
484 Figure L.5 – Histogram for random variable Factor1g1800
Table L.8 – Experimental thresholds to have a 95 % probability that the maximum measured SAR value from the area scan will also have a peak spatial-average SAR
485 L.3.3 Test reduction applicability example
Table L.9 – SAR values from the area scan (GSM 900 band)
486 L.4 Other statistical approaches to search for the high SAR test conditions
L.4.1 General
L.4.2 Test reductions based on a design of experiments (DOE)
Table L.10 – SAR values from the area scan (GSM 900 band)
487 L.4.3 Analysis of unstructured data
488 Annex M (informative) Applying the head SAR test procedures
Table M.1 – SAR results tables for example test results – GSM 850
489 Table M.2 – SAR results table for example test results – GSM 900
Table M.3 – SAR results table for example test results – GSM 1800
490 Table M.4 – SAR results table for example test results – GSM 1900
491 Annex N (informative) Studies for potential hand effects on head SAR
N.1 Overview
N.2 Background
N.2.1 General
492 N.2.2 Hand phantoms
N.3 Summary of experimental studies
N.3.1 General
N.3.2 Experimental studies using fully compliant SAR measurement systems
N.3.3 Experimental studies using other SAR measurement systems
493 N.4 Summary of computational studies
N.5 Conclusions
494 Annex O (informative) Quick start guide
495 Figure O.1 – Quick guide flow-chart
496 Table O.1 – Quick start guide: SAR evaluation steps
498 Bibliography
BS EN 62209-1:2016 - TC
$280.87